"Conformance of a 3T radiotherapy MRI scanner to the QIBA Diffusion Pro" by Madeline E. Carr, Kathryn E. Keenan et al.
Purpose: To assess the technical performance of the apparent diffusion coefficient (ADC) on a dedicated 3T radiotherapy scanner, using a standardized phantom and sequences. Investigations into factors that could impact the technical performance of ADC in the clinic were also completed, including changing the slice-encoded imaging direction and the reference sample ADC value. Methods: ADC acquisitions were performed monthly on an isotropic diffusion phantom over 1 year. Measurements of ADC %bias, coefficients of variation for short-/long-term repeatability and precision (CVST/CVLT and CVP), and...